Monaco Radiation Treatment Planning Workstation, Monaco Release 2.03.00 and above To create treatment plans for any cancer patient for whom external beam intensity modulated radiation therapy (IMRT) has been prescribed.

This recall has been terminated (originally issued June 20, 2011).

Company
Computerized Medical Systems Inc
Recall Initiated
January 13, 2011
Posted
June 20, 2011
Terminated
April 2, 2014
Recall Number
Z-2598-2011
Quantity
137
Firm Location
Maryland Heights, MO
Official Source
View on FDA website ↗

Reason for Recall

Monaco: When the isocenter location is edited in the IMRT Activity screen, the shift coordinates are not updated in the IMRT Plan report.

Distribution

Worldwide Distribution - USA (Nationwide distribution to AK, AZ, CA, CO, FL, IL, IN, MA, MI, MN, MO, NC, NM, NY, OH, OR, PA, SC, TX, VA, WA, and WI and the countries of Algeria, Australia, Austria, Belarus, Belgium, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, India, Israel, Italy, Japan, Mexico, Netherlands, New Zealand, Phillipines, Poland, Portugal, Singapore, Slovenia, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, and the United Kindom.

Lot / Code Info

Monaco Release 2.03.00and above

Root Cause

Software design

Action Taken

Computerized Medical Systems Inc drafted a User Notice which explains the reason for correction, the clinical impact, and the workaround method. The User Notice was sent January 13, 2011. A return postcard is included for the customer to confirm receipt of the User Notice. The firm intends for this issue to be corrected in Monaco Release 3.00 in March 2011.