RecallDepth

2.0/2.4 Cannulated Screw Driver Bit, part #210-24-003

This recall has been terminated (originally issued April 26, 2010).

Company
Trilliant Surgical Ltd.
Recall Initiated
April 14, 2009
Posted
April 26, 2010
Terminated
May 3, 2010
Recall Number
Z-1438-2010
Quantity
32 units
Firm Location
Houston, TX
Official Source
View on FDA website ↗

Reason for Recall

Firm received several complaints of field failures, e.g., "broke during surgery" and the firm initiated a design change. A communication to sales staff 04/14/2009 instructed them to recover and reconcile these products from user physicians.

Distribution

WI, TX, MO, NM, MN, OK, MA, CA, GA, ND, NE, IA, NY, LA, and MD

Lot / Code Info

Lot #0924081

Root Cause

Other

Action Taken

Firm notified sales force by e-mail on 4/14/2009 of the need to recover and reconcile this product . Additional documentation is not available from the firm. Questions may be directed directly to the firm at 800-495-2919.

More recalls by Trilliant Surgical Ltd.

View all recalls by this company →