Radial Assist RAD BOARD, used to support the weight of a patient's arm and supplies for a medical procedure. The Merit RAD BOARD is a rigid and stiff body board intended for use for various medical purposes. RAD BOARD was specifically designed to support the weight of a patient's arm and supplies for medical procedure, in order to have optimal access to upper extremity vasculature, including radial and brachial arterial and venous access. The RAD BOARD is partially lined with a layer of lead-free Xenolite TB for additional radiation scatter protection.

Company
Merit Medical Systems, Inc
Recall Initiated
December 18, 2013
Posted
January 30, 2014
Terminated
February 13, 2014
Recall Number
Z-0901-2014
Quantity
11
Firm Location
South Jordan, UT

Reason for Recall

Merit Medical Systems, Inc. is voluntarily recalling one lot (B507171) of RAD BOARD RB 100 devices. The affected devices are missing the main label which graphically depicts patient orientation on the board. The issue may allow the board to be oriented in the wrong direction where the semi-radiopaque material in the board (embedded Xenolite) could obscure patient anatomy during fluoroscopy. There

Distribution

Worldwide Distribution - USA including NJ, TX, WV, MN, GA, and Internationally to Mexico and Japan.

Lot / Code Info

lot B507171

Root Cause

Labeling Change Control

Action Taken

Merit Medical Systems, Inc.sent a letter voluntarily recalling one lot (B507171) of RAD BOARD" RB100 devices. The affected devices are missing the main label which graphically depicts patient orientation on the board. The issue may allow the board to be oriented in the wrong direction where the semi-radiopaque material in the board (embedded Xenolite) could obscure patient anatomy during fluoroscopy. There have been no reports of patient harm or injury from Merit customers as a result of this issue. Merit has chosen to remove this lot from the field by replacing the unlabeled RAD BOARDs. Merit requests that customers stop using these devices immediately until the product has been replaced by their Merit sales representative. Customers with questions were instructed to call 1-801-208-4344 or 1-801-208-4365. For questions regarding this recall call 801-208-4301.