Merit Maestro Microcatheter, 2.4F 130cm Straight, REF28MC24130ST, Rx Only, Non-pyrogenic Sterile if package is unopened or undamaged.

Company
Merit Medical Systems, Inc
Recall Initiated
September 26, 2014
Posted
October 31, 2014
Terminated
November 19, 2014
Recall Number
Z-0162-2015
Quantity
56 units
Firm Location
South Jordan, UT

Reason for Recall

These microcatheters were packaged with a 2.8 F distal tip, rather than the labeled 2.4 F.

Distribution

Nationwide and to India. No govt/military/VA consignees.

Lot / Code Info

Catalog No: 28MC24130ST, Lot H665694, H669824, Exp 2017 and Catalog No: 28MC24150SN, Lot H665696, Exp 7/31/2014

Root Cause

Mixed-up of materials/components

Action Taken

Consignees were notified via letter on September 26, 2014. Written notification included product identification information, instruction to immediately quarantine any devices and discontinue use, ensure all personnel to whom devices were distributed are made aware of this field action, and instructions to contact their Merit representative. An email notification was sent to US Merit sales representatives with accounts that had received affected product. The affected sales representatives were provided a Customer Response Form applicable to their account, a representative copy of communications sent directly to their affected accounts, and a copy of the Frequently Asked Questions. Merit sales representatives were instructed to visit the account and assist with the return of the affected units back to the Merit Field Assurance Department.