Boston Scientific TSX Transseptal Needle, Model Number M004TSX60o; Used to create the primary puncture in the interatrial septum when passing an intruder and/or catheter through the septum from the right side of the heart to the left side.

Company
Merit Medical Systems, Inc
Recall Initiated
March 15, 2016
Terminated
August 22, 2016
Recall Number
Z-1375-2016
Quantity
25
Firm Location
South Jordan, UT

Reason for Recall

Merit Medical Systems, Inc. is voluntarily conducting a recall of one lot of the TSX" Transseptal Needle due to a discrepancy between the labeling and the actual configuration of the device in the packaging. The curve dimensional values on the labels of this lot incorrectly indicate a 50 curve, rather than the correct 86 curve

Distribution

Distributed to one customer in the US.

Lot / Code Info

Lot Q823423, Exp 7/31/2018

Root Cause

Software design

Action Taken

The consignee was provided verbal and written notification of the field action on March 15, 2016. There was a total of one (1) consignee, an OEM customer in the USA, who received 25 units. Written notification included product identification information, instruction to immediately quarantine any devices and discontinue distribution, and instructions to initiate a product recall in accordance with their company's established policies for customers to whom they shipped the affected Merit product. An email notification was sent to the Merit sales representative of the affected account. The sales representative was provided a Customer Response Form applicable to their account. The sales representative was instructed to assist the account with the completion of the Customer Response Form and the return of the affected units back to the Merit Field Assurance Department.

More recalls by Merit Medical Systems, Inc

View all recalls by this company →