IMPAX CV Results Manager/Results Manager Administration Tool The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting. The suite is comprised of modular software that runs on industry standard personal computers and servers, utilizing Windows 2000/2003/XP operating systems. Image data storage/display utilizes DICO< (Digital Imaging and Communications in Medicine) data primarily from independent medical devices.

Company
Agfa Corp.
Recall Initiated
April 15, 2009
Posted
February 25, 2011
Terminated
October 13, 2011
Recall Number
Z-1435-2011
Quantity
6 devices
Firm Location
Greenville, SC

Reason for Recall

User could modify a measurement's units and that change would not be reflected in echocardiography formulas used within the Adult echocardiography RM.

Distribution

Nationwide distribution: USA including states of: IL, MT, NY, OR and SC.

Lot / Code Info

Software Versions: RM 2.03, 2.04 & 2.06

Root Cause

Other

Action Taken

The firm, Agfa Healthcare, sent individual letters dating from April 15, 2009 to August 26, 2009, to the customers. The letters described the product, problem and actions to be taken. Agfa provided the affected customer with a complete list of their impacted patients and the measurements. Agfa also scheduled unit corrections with their customers and received acknowledgement of the letters at the time of scheduling the unit corrections. Service Bulletin DIS026.09E was issued January 2009 to inform users that Agfa will no longer allow customers to request unit changes for measurements that have formulas. Note: Technical Support ran diagnostic scripts for each customer and documented in GSC cases for each consignee/customer. If you have any questions, contact Agfa Healthcare at (949) 830-4959.

More recalls by Agfa Corp.

View all recalls by this company →