Prelude Short Sheath Introducer, 7F-SMT, 4 cm, REF No: PSS-7F-4-038MT, Sterile EO, Merit Medical, South Jordan, Utah. Intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. Can also provide access to a native or synthetic graft used for hemodialysis.

Company
Merit Medical Systems, Inc
Recall Initiated
December 17, 2010
Posted
January 28, 2011
Terminated
July 11, 2011
Recall Number
Z-0941-2011
Quantity
378 units
Firm Location
South Jordan, UT

Reason for Recall

Introducer tips may detach and embolize during use, or the tip material may elicit a thrombogenic response.

Distribution

Nationwide Distribution

Lot / Code Info

Lot Number: H179575

Root Cause

Component change control

Action Taken

Merit Medical Systems, Inc., issued an "Urgent Product Recall Notice" dated December 17, 2010, and contacted all customers (sales reps and authorized distributors) via email, phone and/or fax. The notice described the product, problem and action to be taken by the customers (Sales reps/Distributors). The customers were instructed to contact all their customers to identify, quarantine and return all unused inventory and extend the recall to any of their customers who may have received the affected products. The customers were instructed to complete and return the Manufacturer RECALL NOTIFICATION and Product Retrieval Form via fax to 1-801-208-3378 and send original to: the Manager, Customer Service, Merit Medical Systems, Inc., 1600 West Merit Parkway, South Jordan, UT 84095-2416. Questions should be directed to the Senior Product Manager at 801-208-4344 or the Manager, Customer Service at 801-208-4365.

More recalls by Merit Medical Systems, Inc

View all recalls by this company →