Xoran MiniCAT for ENT, Xoran Technologies, Incorporated, Ann Arbor, MI 48103. This device is a cone beam CT scan for sinue, temporal bones and maxillofacial imaging.

This recall has been terminated (originally issued November 27, 2007).

Company
Xoran Technologies, Inc.
Recall Initiated
July 13, 2006
Posted
November 27, 2007
Terminated
March 13, 2008
Recall Number
Z-0296-2008
Quantity
5
Firm Location
Ann Arbor, MI
Official Source
View on FDA website ↗

Reason for Recall

The hard drive may crash, resulting in loss of data from patient CT scans.

Distribution

Nationwide Distribution: California, South Dakota, Tennessee, Utah and West Virginia.

Lot / Code Info

Serial Numbers: ENT00009, ENT00051, ENT00084, ENT00086 and ENT00088.

Root Cause

Finished device change control

Action Taken

Consignee accounts were visited between 7/25/06 and 8/10/07 and the hard drives on their systems were replaced.