Stratus CS STAT Fluorometric Analyzer BHCG TestPak (CBHCG) in vitro diagnostic.

Company
Dade Behring Inc.
Recall Initiated
April 3, 2006
Posted
May 19, 2006
Terminated
August 7, 2006
Recall Number
Z-0874-06
Quantity
126 cartons of 100 TestPaks
Firm Location
Newark, DE

Reason for Recall

Foil sealing of theTestPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.

Distribution

Nationwide to hospitals and government accounts in addition to International accounts in Canada, Saudi Arabia, Bangladesh, UAE, India, Japan, Denmark, and Australia.

Lot / Code Info

Lot #446003002 and 446058002.

Root Cause

Other

Action Taken

Customers were provided clear written instructions (04/06/2006) to discontinue use and discard remaining inventory of products over 2 months old. In addition, each customer is being contacted by phone as an effectiveness verification. Recalling firm is also instructing Stratus CS customers to process all samples and dilutions in duplicate.