EasyLink Informatics System, software version 3.0, Part Number 1000034941, DADE BEHRING INC., Newark, DE.
- Company
- Dade Behring Inc.
- Recall Initiated
- July 3, 2007
- Posted
- August 22, 2007
- Terminated
- November 7, 2007
- Recall Number
- Z-1174-2007
- Quantity
- 72 units
- Firm Location
- Newark, DE
Reason for Recall
Sample ID and Patient results could be associated with the wrong patient
Distribution
Worldwide: USA including states of AZ, CA, FL, IL, MA. MI, MO, NC, OH, PA, TX, and WV. The product was also shipped to other Dade facilities in MO, DE, IL, CT, and Belgium.
Lot / Code Info
Part number 1000034941
Root Cause
Other
Action Taken
The recalling firm issued a Field Corrective Action letter dated 7/3/07 to inform their customers of the problem and that a software patch would be sent for them to download. The recalling firm plans to telephone the customers after the patch has been released to verify that it has been installed.