Dimension EXL Clinical Chemistry System used with software versions 8.5.1 and 8.5.1SP3. The system is a discrete, random-access, microprocessor-controlled, integrated instrument/chemistry system that measures a variety of analytes, including enzyme activities, in body fluids for in vitro diagnostic use.

Company
Dade Behring Inc.
Recall Initiated
September 30, 2008
Posted
January 13, 2009
Terminated
May 5, 2009
Recall Number
Z-0581-2009
Quantity
57 units
Firm Location
Newark, DE

Reason for Recall

Error messages, "Wash Aspirator Failure" and "Wash station unable to recover" are not generated when one or both HM wash probes are totally occluded.

Distribution

Nationwide Distribution including states of AR, CA, CO, IA, ID, IL, IN, LA, MD, MI, MO, MS, NC, NE, OH, OK, PA, SD, TN, TX, VA, and WI.

Lot / Code Info

Software versions 8.5.1 and 8.5.1SP3.

Root Cause

Software design

Action Taken

The recalling firm notified consignees by telephone to inform them of the problem. A follow-up letter, "Urgent Field Safety Notice" dated 9/30/08 was issued with information describing the problem, the possible impact of an occluded wash probe on assay results, and emphasizing regular performance of routine wash probe maintenance as detailed in the operator guide to prevent wash probe occlusion. For questions, please contact the Siemens Technical Solution Center at 800-441-9250.