SonoCalc IMT 3.0, a windows based software used in conjunction with images from high-resolution ultrasound systems (MicroMaxx, TITAN, and SonoSite180PLUS). This software generates a report with the patients intima media thickness (IMT) based on average thickness of the carotid artery.

This recall has been terminated (originally issued April 6, 2006).

Company
Sonosite, Inc.
Recall Initiated
October 26, 2005
Posted
April 6, 2006
Terminated
April 10, 2006
Recall Number
Z-0713-06
Quantity
155 units of the SonoCalc product to about 80 customers/distributors
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

When SonoCalc IMT 3.0 software is used with the SonoSite 180Plus ultrasound system, the software miscalculates the intima-media-thickness, a measurement of risk for cardiovascular or cerebrovascular events. Underestimates percentage of stenosis, false negative result.

Distribution

Nationwide, Kuwait, Mexico, South Africa, Chile, Spain, Japan, Puerto Rico, Australia. A list of 155 cases in which the software was sent out lists medical facilities, physicans, and SonoSite Distributors.

Lot / Code Info

All SonoCalc IMT version 3.0

Root Cause

Other

Action Taken

On October 26, 2005, the company sent their customers a User Guide Addendum that told customers how to verify that the image calibration data was correct.