SLT transducer is used on the SonoSite MicroMaxx ultrasound system. The SLT Transducer label is labeled in part: " MicroMaxx SLT...Manufactured for SonoSite, Inc., Bothell, WA 98021".

This recall has been terminated (originally issued October 1, 2010).

Company
Sonosite, Inc.
Recall Initiated
November 2, 2007
Posted
October 1, 2010
Terminated
October 5, 2010
Recall Number
Z-2655-2010
Quantity
38 units
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

Sterilization of the SonoSite SLT ultrasound surgical transducer with the Sterrad NX System may result in damage to the transducer.

Distribution

Devices were distributed to medical centers and hospitals in AR, CA, IN, KS, MA, MD, MI, NC, NE, NJ, NV, OH, SC, TN, TX, WI, Washington DC and to FOREIGN COUNTRIES Australia, China, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Korea, Republic of, Libyan Arab Jamahiriya, Mexico, Spain, United Arab Emirates, United Kingdom, and United States.

Lot / Code Info

38 units are identified with Serial numbers:  0361MW, 036MH5, 0365TR, 036MH3, 0360VT, 036MGZ, 03661X, 035TCZ, 037Y9N, 036MH1, 0365TN, 0360VW, 03661W, 035KXD, 037Y9M, 035KXC, 037Y84, 036MH6, 037Y9L, 0365TX, 03616X, 037Y8D, 037Y83, 0365TW, 035L1M, 036MH8, 035TXC, 0360TN, 0361MQ, 03616Z, 035NTF, 034RZG, 037YKG, 037YKH, 0361CG, 036MH9, 0365TT and 037YKM.

Root Cause

Process control

Action Taken

On November 2, 2007, SonoSite sent out the Important Information Regarding Sterrad NX Incompatibility letter and the Technical Bulletin 005/07 Rev A to their consignees. The letter advises the consignees that SonoSite has recently found that sterilization of the SonoSite SLT and LAP surgical ultrasound transducers with the Sterrad NX system may result in damage to the transducer. The Technical Bulletin 005/07 Rev B dated November 2, 2007 informs the consignees that the damage has been observed primarily on the circuit board in the transducer connector, and cable and laparoscope damage may also be occurring. All damages have occurred after processing using the Sterrad NX systems in its "Advanced" cycle and the potential damage from the use of "normal" cycle is being evaluated by the firm. Consignees should call the Technical Support at 1-425-951-1330 or 1-877-657-8118 or e-mail to service@sonosite.com for any questions regarding the recall.