P21x/5-1 MHz transducer, Part Number P07698-23 Product Usage: The P21x/5-1 MHz phased array transducer is designed for cardiac, abdominal and obstetric, TCD and orbital exams. It supports 2D with SonoMB Multi Beam Technology and Auto Gain capabilities, SonoHD Imaging Technology, Tissue Harmonic Imaging, M-Mode, Velocity Color Flow Doppler, Color Power Doppler (CDP), Pulsed Wave (PW), TDI Pulsed Wave, and Continuous Wave Doppler.

This recall has been terminated (originally issued October 18, 2013).

Company
Sonosite, Inc.
Recall Initiated
September 11, 2013
Posted
October 18, 2013
Terminated
December 11, 2013
Recall Number
Z-0027-2014
Quantity
10 units
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

FUJIFILM SonoSite Inc. remove P21x/5-1 Transducer because the firm discovered the Biopsy Bracket Sensor is not working as intended.

Distribution

Worldwide Distribution - USA Nationwide in the states of Florida, Georgia, Oklahoma, Pennsylvania, Tennessee, Texas, West Virginia, and in the country Japan and Italy.

Lot / Code Info

Serial Number: 03T861; 03T863; 03T865; 03T867; 03T869; 03T85Z; 03T86C; 03T86F; 03T86H; 03T86K.

Root Cause

Error in labeling

Action Taken

SonoSite started calling their consignees on September 13, 2013. For customers who could not be reached via phone, a letter Urgent -Medical Device Removal will be sent on September 17, 2013. Affected customers will be offered a replacement transducer (P07698-24) free of charge. Consignees in the U.S. can call 1-877-657-8118, Monday-Friday, 6:00 a.m. - 5:00 p.m. (PST) or send E-mail to service@sonosite.com. Customers outside the US can call 425-951-1200.