Misys Laboratory version 6.1 - Microbiology Module
This recall has been terminated (originally issued October 22, 2005).
- Company
- Misys Healthcare Systems
- Recall Initiated
- September 29, 2005
- Posted
- October 22, 2005
- Terminated
- February 10, 2006
- Recall Number
- Z-0088-06
- Quantity
- 67
- Firm Location
- Tucson, AZ
- Official Source
- View on FDA website ↗
Reason for Recall
Customer complaint during testing of module prior to production use. Software problem when the user has indicated to enter a coded comment instead of using the defined keyboard, then enters 100,000, the system willfile the number 100 instead of number 100,000.
Distribution
Nationwide and Canada, Denmark, Emirates, Saudi Arabia, United Kingdom
Lot / Code Info
version 6.1
Root Cause
Other
Action Taken
Telephone calls and emails dated 9/29/05 to all consignees in production. Product Safety Noticed dated 9/29/2005 was faxed to all consignees testing the product.