RecallDepth

Misys Laboratory with Lab Access Results application for versions 5.3 and later.

This recall has been terminated (originally issued June 28, 2005).

Company
Misys Healthcare Systems
Recall Initiated
May 13, 2005
Posted
June 28, 2005
Terminated
February 10, 2006
Recall Number
Z-0938-05
Quantity
258
Firm Location
Tucson, AZ
Official Source
View on FDA website ↗

Reason for Recall

Software anomaly may rarely allow modification of previously reported specimen results.

Distribution

Nationwide and United Kingdom, Ireland, Denmark, Canada, Saudi Arabia, United Arab Emirates

Lot / Code Info

Versions 5.3 and later.

Root Cause

Other

Action Taken

Product Safety Notice PSN-05-01 was sent by fax on May 13, 2005 and includes a procedural work-around to avoid the defect until the firm is able to provide a software patch.