Mysis Laboratory version 6.1
This recall has been terminated (originally issued June 28, 2005).
- Company
- Misys Healthcare Systems
- Recall Initiated
- May 10, 2005
- Posted
- June 28, 2005
- Terminated
- August 30, 2005
- Recall Number
- Z-0939-05
- Quantity
- 21
- Firm Location
- Tucson, AZ
- Official Source
- View on FDA website ↗
Reason for Recall
When a previously reported microbiology organism result is deleted on a selected patient, reported susceptibilities will not stay with associated organisms and one or more suceptibilities on the culture will be deleted.
Distribution
Nationwide and United Kingdom and Denmark
Lot / Code Info
Version 6.1
Root Cause
Other
Action Taken
Firm sent Product Safety Notice PSN 05-02 on May 10, 2005 which also provides a procedural work-around and an invitation to obtain a service pack to correct the software flaw.