MicroMaxx LAP/12-5 MHz Transducer, Laparoscopic Ultrasound Transducer

This recall has been terminated (originally issued March 29, 2007).

Company
Sonosite, Inc.
Recall Initiated
February 5, 2007
Posted
March 29, 2007
Terminated
June 1, 2007
Recall Number
Z-0661-2007
Quantity
6 Devices
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

Laparoscopic transducer exceeds the limits specified for radio-frequency emissions by up to 10db.

Distribution

Distribution to USA (Tennessee & Ohio), France, Korea, and Germany.

Lot / Code Info

Serial Numbers: 037XT6, 0379P5, 037WMJ, 0379P1, 0379P2, 0379P0 Part Number: P05507-10

Root Cause

Other

Action Taken

On 2/5/2007 firm sent by FED-Ex notification letters to firm representatives. Representatives began hand delivery of recall letter to consignees and retrieval of LAP transducer for return delivery to firm. The letter advised customers that the LAP transducer exceeded the limits specified for radio-frequency emissions. The firm also advised that exceeding the limit may cause the LAP transducer to interfere with surrounding equipment that may be susceptible to radiated emissions and cause surrounding equipment to not perform as expected.