Abbott Latex-Free Primary I.V. Plumset Convertible Pin, 95 inch with Injection Site and Secure Lock, No. 19020; Made in Costa Rica, Abbott Laboratories, North Chicago, IL 60064
This recall has been terminated (originally issued June 17, 2003).
- Company
- Abbott Laboratories Hpd/Add
- Recall Initiated
- May 13, 2003
- Posted
- June 17, 2003
- Terminated
- November 20, 2003
- Recall Number
- Z-0936-03
- Quantity
- 4,656 units
- Firm Location
- Abbott Park, IL
- Official Source
- View on FDA website ↗
Reason for Recall
The I.V. administration set incorrectly lists the priming volume from the cassette inlet to distal male adapter as 13.5 mL instead of 11.3 mL
Distribution
Miami, FL
Lot / Code Info
List number 19020-04-12, lot 941845H
Root Cause
Other
Action Taken
Recall letter dated 5/12/03 sent to the account on 5/13/03, informing them of the labeling discrepancy and requesting the return of any inventory of the affected lot to Abbott for credit.