Hitachi CXR-4 CT Computed Tomography X-ray System (including SceptreP3 PET/CT)

This recall has been terminated (originally issued June 3, 2006).

Company
Hitachi Medical Systems America Inc
Recall Initiated
February 21, 2006
Posted
June 3, 2006
Terminated
November 13, 2008
Recall Number
Z-1078-06
Quantity
15 units
Firm Location
Twinsburg, OH
Official Source
View on FDA website ↗

Reason for Recall

The device has a software anomaly which affects the system's ability to create Multiplanar Reconstructions (MPR). Some images from the resulting series do not display correctly, or display at all, when MPR was run on them.

Distribution

The recalled MRI systems were distributed throughout the continental United States.

Lot / Code Info

Serial Number: CXR46201-46213 PC 46002-36011

Root Cause

Other

Action Taken

Hitachi Medical Systems America, Inc. became aware of an intermittent problem in the Multi-planar Reconstruction process when covering long ranges that could result in blank frames in the displayed reconstructions. HMSA also discovered a second software-related problem with the 50mm scale displayed on filmed images. In cases where the image has been magnified before filming, the 50mm scale is not correctly sized to the magnified image. Software corrections have already been installed on all units in order to correct the first of these problems (MPRs). According to the firm''s Device Notification Letter issued on 2/21/2006, a second software upgrade [which the firm states address both issues] will be released on or before 2/27/2006 and this correction will be performed by HMSA''s service personnel during each customer''s next scheduled PM. The Device Notification Letter also instructs the customers to call their Field Service Representatives or the HMSA Applications Helpline if they have any questions about the notice, or if any additional problems are encountered.