Clinical Chemistry Uric Acid; LN 7D76-20
This recall has been terminated (originally issued July 20, 2004).
- Recall Initiated
- March 18, 2004
- Posted
- July 20, 2004
- Terminated
- February 10, 2012
- Recall Number
- Z-0769-04
- Quantity
- 796 kits.
- Firm Location
- South Pasadena, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Unacceptable upward drift of results in control runs within 24 hour period.
Distribution
Nationwide and Mexico, Equador, Puerto Rico, Hong Kong, Bahamas, Venezuela, Argentina, Canada, New Zealand, Columbia, Brazil, England, Trinidad.
Lot / Code Info
Lot number: 95015HW00
Root Cause
Other
Action Taken
Firm sent recall letters 3/18/2004 suggesting destruction and/or running QC tests every eight hours instead of every 24 hours as is in labeling.