One Touch Ultra Test Strips, part number 020-994-02, manufactured by LifeScan Inc, Milpitas, CA, 25 count. Intended to be used for quantitative measurement of glucose in fresh capillary whole blood. Intended for use outside the body (in vitro diagnostic use) by people with diabetes at home and healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. Intended for use on the finger, arm, or palm.

Company
Lifescan Inc
Recall Initiated
September 19, 2008
Posted
November 21, 2008
Terminated
December 27, 2010
Recall Number
Z-0004-2009
Quantity
35308 vials
Firm Location
Milpitas, CA

Reason for Recall

Products exceed inaccuracy threshold , with many complaints resulting from inaccurate low results.

Distribution

Nationwide Distribution --- including Puerto Rico.

Lot / Code Info

Lot number 2829235

Root Cause

Process control

Action Taken

Urgent: Medical Device Recall letters, dated 09/19/2008, were sent to consumer end users, direct account distributors and wholesalers, health care providers, and pharmacists notifying them of the recall. The letter states that if customers have strips from Lot 2829235, they should call LifeScan at 1-866-247-1281 and tell customer service that they have strips that need to be replaced. Customer service and QA scripts were provided to LifeScan staff to provide to any callers who inquire about the recall. The distributors and wholesalers were to locate and sequester any inventory from the affected Lot 2829235; request all recalled inventory from customers; call Carolina Logistics for an RGA; and when only the affected returned product is received, a credit will be issued.

More recalls by Lifescan Inc

View all recalls by this company →