Medline EndurO2 Oxygen Concentrator with Oxygen Monitor; Portable oxygen generator; Reorder Number: HCS02; Packaged in 1 unit per case; Product is an oxygen generator designed to provide oxygen to patient who has been prescribed oxygen therapy by a physician.
- Company
- Medline Industries Inc
- Recall Initiated
- March 4, 2008
- Posted
- July 23, 2008
- Terminated
- June 17, 2010
- Recall Number
- Z-1473-2008
- Firm Location
- Mundelein, IL
Reason for Recall
Overheating of Components: There is the potential for an electrical component in the oxygen concentrator to overheat, causing damage to the unit and making it nonfunctional.
Distribution
Nationwide Distribution
Lot / Code Info
Reorder Number: HCS02; All units produced since July 2006.
Root Cause
Device Design
Action Taken
Consignees were notified by a letter sent on 3/4/08. The letter informed users that the recalling firm will replace the affected units with new units or issue a credit for the returned recalled units. Distributors were requested to notify their accounts of the recall. The letter requested users to complete and return an enclosed response sheet, indicating the number of units to be replaced or returned for credit. For additional information, contact 866-359-1704.