RecallDepth

Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.

Company
Medline Industries Inc
Recall Initiated
May 11, 2016
Terminated
June 3, 2019
Recall Number
Z-2404-2016
Quantity
903 Cases and 4,090 Sleeves (8,605,000 Medicine Cups)
Firm Location
Mundelein, IL

Reason for Recall

There is 5 mL printed next to the increment on the medicine cups instead of 10 mL. The cup has the correct mL markings for the other gradations, but sequentially, the markings are listed as 2.5 mL, 5 mL, 7.5 mL, and 5 mL.

Distribution

Continental U.S.

Lot / Code Info

Lot Numbers 122211502 122212515

Root Cause

Process control

Action Taken

Medline Industries sent a an URGENT RECALL letter dated 5/11/2016 to all affected consignees via certified mail. The URGENT RECALL letter described the reason for the recall and the actions to take.

More recalls by Medline Industries Inc

View all recalls by this company →