RecallDepth

Rigid Suction Canister 3000 CC and 2000 CC.

Company
Medline Industries Inc
Recall Initiated
August 10, 2016
Terminated
June 3, 2019
Recall Number
Z-0352-2017
Quantity
7,680 units
Firm Location
Mundelein, IL

Reason for Recall

Risk of implosion when the Rigid Suction Canister is subjected to excessive negative pressure during use.

Distribution

US Distribution to the following states: CA , CO , FL , GA, GA , IL , MA, MA , MD , MI, MO , NC , NE , NJ , NY, NY , OH, OK , PA, TX and WA.

Lot / Code Info

RIGID CANISTER 2000 cc and 3000 cc with product code OR220 and OR230. Affected Lot Numbers 9961601xxxx through 9961604xxxx

Root Cause

Process control

Action Taken

Medline Industries, Inc. sent an Urgent Recall Immediate Action Required letter dated August, 23, 2016, and destruction form via first class mail to affected customers on August 17, 2016 and August 23, 2016. The firm recommended the following actions: REQUIRED ACTION: 1. Immediately check your stock for item number OR220 and OR230 with lot numbers beginning with 9961601xxxx through 9961604xxxx (listed on the original case packaging). Destroy all affected products. If the rigid canisters are NOT in the original case packaging, we are instructing you to destroy them and we will send replacements. 2.Please return the completed enclosed destruction form listing the quantity of product(s) destroyed. Even if you do not have any of this product please complete and return the form, as indicated on the destruction form. When we receive your completed destruction form your account will receive replacements. 3.If you are a distributor, promptly notify any of your customers that may have received affected product. If you have any questions, please contact 866-359-1704.

More recalls by Medline Industries Inc

View all recalls by this company →