SonoSite L25 Needle Guide Kit, 21 GAUGE, Part Number P04053-01. Manufactured for: SonoSite, Inc., Bothell, WA 98021 Each kit is packaged in a hermetically sealed clear plastic pouch with white backing. Product is labeled as Sterile. 24 pouches are packaged in a white cardboard box and labeled as SonoSite L25 Needle Guide Kit, 21 GAUGE REF# P04053-01, Quantity: 24, CONTENTS STERILE IF PACKAGE IS INTACT. Mfg for: Sonosite, Inc. 888/482-9449*425/951-1200, Bothell, WA 98021 USA. www.sonosite.com

This recall has been terminated (originally issued January 19, 2006).

Company
Sonosite, Inc.
Recall Initiated
December 12, 2005
Posted
January 19, 2006
Terminated
July 31, 2006
Recall Number
Z-0407-06
Quantity
355 boxes/24 kits per box
Firm Location
Bothell, WA
Official Source
View on FDA website ↗

Reason for Recall

22 gauge needle guides were intermixed with 21 gauge guides and labeled as 21 gauge needle guides.

Distribution

Product was distributed to distributors and hospitals nationwide and internationally.

Lot / Code Info

Box Lot Code: M778940; M779110; M785430; M790190; M791960; M793010; M793020; M793530; M799610; M799620; M803570; M803580; M806400; M806410; M806420; M809020; M810990; M823040; M830430.  Pouch Lot Code: M761640; M778400; M781250; M783140; M785270; M791720; M797960; M801010; M802700; M804630; M807010; M811660; M831060; M832790

Root Cause

Other

Action Taken

Affected customers were contacted via telephone between December 12-19, 2005. Provided instructions for returning defective product. Notification letter was distributed to all customers on December 14, 2005 instructing customers to discontinue use of the needle guide and return remaining portion.