SoftID.NET versions 2.0.0.6 - 2.0.0.19. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. SoftID.NET is to be used in a medical or clinical laboratory setting by knowledgeable, trained, and experienced professionals. SoftID.NET supports bedside collection of patient specimens used in diagnostic testing. This product is integrated with Soft Computer's SoftLab, SoftScape and can also interface with all other SCC laboratory systems and Hospital Information Systems (HIS).

This recall has been terminated (originally issued November 15, 2010).

Company
Scc Soft Computer
Recall Initiated
August 28, 2009
Posted
November 15, 2010
Terminated
November 16, 2011
Recall Number
Z-0360-2011
Quantity
41
Firm Location
Clearwater, FL
Official Source
View on FDA website ↗

Reason for Recall

SCC Soft Computer issued a Correction for SoftID.NET versions 2.0.0.6 - 2.0.0.19. When using the Cancel functionality, if the collection list refreshes during the transaction and changes the position of the patient on the list, the wrong patient order can be cancelled. All affected clients have been notified.

Distribution

Nationwide distribution: AR, CA, FL, IL, IN, KS,LA, MA, MI, MN, MO, NC, ND,NH, NJ, NY, OH, PA, RI, TN, TX, VA and WI.

Lot / Code Info

Versions: SoftID.NET 2.0.0.6 - 4/3/2007,  SoftID.NET 2.0.0.7 - 6/21/2007,  SoftID.NET 2.0.0.8 - 8/29/2007,  SoftID.NET 2.0.0.9 - 10/01/2007,  SoftID.NET 2.0.0.10 - 10/25/2007,  SoftID.NET 2.0.0.11 - 11/07/2007,  SoftID.NET 2.0.0.12 - 2/08/2008,  SoftID.NET 2.0.0.13 - 5/13/2008,  SoftID.NET 2.0.0.14 - 6/12/2008,  SoftID.NET 2.0.0.15 - 8/15/2008,  SoftID.NET 2.0.0.16 - 10/15/2008,  SoftID.NET 2.0.0.17 - 12/15/2008,  SoftID.NET 2.0.0.18 - 2/02/2009,  SoftID.NET 2.0.0.19 - 5/18/2009.

Root Cause

Software design

Action Taken

SCC Soft Computer notified Customers on August 28, 2009, via the proprietary communication software entitled "Risk-to-Health Notification Task Verbiage" dated August 17, 2009. The notification described the product, problem and action to be taken by customer. The customers were instructed to acknowledge receipt of the notification; document a request to load the software, if they wish to receive the correction; document that the Alternative Working Solution will be used by the customers facility until the correction has been taken live. If you have any questions, contact (727) 789-0100.