RecallDepth

Medical convenience kits UDPG88W PICC G-TUBE PACK-230209

This recall is currently active, issued October 16, 2025. It was issued by American Contract Systems Inc..

Company
American Contract Systems Inc.
Recall Initiated
September 4, 2025
Posted
October 16, 2025
Recall Number
Z-0186-2026
Quantity
32,433 total
Firm Location
Grimes, IA
Official Source
View on FDA website ↗

Reason for Recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Distribution

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Lot / Code Info

UDPG88W UDI-DI 191072216359 lot 7935611 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected

Root Cause

Nonconforming Material/Component

Action Taken

Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.