Medical convenience kits Item Number/Description EVLH67F LEFT HEART CATH PACK FDCL22R CATH LAB ANGIO IHCC03X CATH CARDIAC PACK TNCC02AD CARDIAC CATH TRINITY UICC27AG CARDIAC CATH ANGIO PACK (PS 923077) UIRD89AH RADIAL PACK (PS 56573)
This recall is currently active, issued October 16, 2025. It was issued by American Contract Systems Inc..
- Company
- American Contract Systems Inc.
- Recall Initiated
- September 4, 2025
- Posted
- October 16, 2025
- Recall Number
- Z-0175-2026
- Quantity
- 32,433 total
- Firm Location
- Grimes, IA
- Official Source
- View on FDA website ↗
Reason for Recall
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Lot / Code Info
AMCL05P UDI-DI 191072207296 Lot 7990011 Bag serial no. N/A Lot 8170111 Bag serial numbers 68194152 68194153 68194160 68194176 68194179 68194182 68194190 Lot 8526511 Bag serial numbers 68873157 68874410; EVLH67F UDI-DI 191072232380 Lot 8211011 Bag serial number N/A Lot 8367311 Bag serial number 68570438; FDCL22R UDI-DI 191072185259 Lot 8048511 Bag serial numbers 68045373 68045376; IHCC03X UDI-DI 191072226273 Lot 80111 Bag serial number N/A Lot 8085711 Bag serial numbers 47506640 47506761; TNCC02AD UDI-DI 191072212757 Lot 7970911 Bag serial number N/A Lot 8045911 Bag serial number 47503192 Lot 8426711 Bag serial number 47886615; UICC27AG UDI-DI 191072231307 Lot 8085511 Bag serial number N/A Lot 8167611 Bag serial number 68194108; UIRD89AH UDI-DI 191072231314 Lot 8068111 Bag serial number 47499672 47499908 * if Bag serial number is "n/a", then all of that lot is affected
Root Cause
Nonconforming Material/Component
Action Taken
Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.