Medical convenience kits Item Number/Description FHCP08Y C SECTION PK HICS50I C-SECTION PACK HSCB90B C-SECTION BASIN PACK
This recall is currently active, issued October 16, 2025. It was issued by American Contract Systems Inc..
- Company
- American Contract Systems Inc.
- Recall Initiated
- September 4, 2025
- Posted
- October 16, 2025
- Recall Number
- Z-0181-2026
- Quantity
- 32,433 total
- Firm Location
- Grimes, IA
- Official Source
- View on FDA website ↗
Reason for Recall
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Lot / Code Info
FHCP08Y UDI-DI 191072232724 Lot 8216211 Bag serial number 68258296; HICS50I UDI-DI 191072212351 Lot 8161711 Bag serial number 68273637; HSCB90B UDI-DI 191072141606 Lot 8083511 Bag serial numbers 47506618 47508489 * If bag serial number is listed as "N/A", then whole lot is affected
Root Cause
Nonconforming Material/Component
Action Taken
Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.