Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)

Company
Maquet Medical Systems, Usa
Recall Initiated
October 12, 2023
Posted
November 9, 2023
Recall Number
Z-0290-2024
Quantity
1022 units
Firm Location
Wayne, NJ

Reason for Recall

It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.

Distribution

US Nationwide.

Lot / Code Info

UDI-DI: 04058863074863; Serial Numbers: 90414355 through 90414868

Root Cause

Device Design

Action Taken

An URGENT MEDICAL DEVICE Correction notification letter dated 10/12/23 was sent to customers. Actions to be taken by the Customer: Our records indicate that you have received a CARDIOHELP-i System having a serial number that is affected by this voluntary Medical Device Correction. " Please examine your inventory immediately to determine if you have any affected CARDIOHELP System(s) in your inventory. " Should you have any affected CARDIOHELP-i System(s), your unit(s) will need to be returned to Getinge for correction. Getinge will contact your facility to arrange the return of your unit for correction. A CARDIOHELP loaner can be provided if needed. " Please do not to fixate any accessory at the slide rail until the slide rail holders have been replaced. " Affected CARDIOHELP-i can be used for intra-hospital transport only when placed on a Sprinter cart. " Inter-hospital transport is not permitted until the slide rail holders have been replaced. " Please forward this information to all current and potential CARDIOHELP-i users within your hospital / facility. " If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. " Whether or not your facility has affected product(s) listed in this notice, please complete and sign the attached MEDICAL DEVICE- CORRECTION RESPONSE FORM (Page 4) to acknowledge that you have received this notification. Return the completed form to Getinge by e-mailing a scanned copy to Cardiohelpsliderail2023.act@getinge.com or by faxing the form to 1-866-313-4049. Type of Action by Getinge: A Getinge representative will contact you to arrange for the repair of your CARDIOHELP -i unit(s). Getinge will correct your CARDIOHELP -i unit(s) upon receipt and will return your unit(s) once correction has been completed. This voluntary correction only affects the products listed on Page 1; no other products are affected by this voluntary correction. We apologize for

More recalls by Maquet Medical Systems, Usa

View all recalls by this company →