RecallDepth

Carry bar which is an accessory to a lift used for patients in hospitals and other facilities. Recall impacts the following model numbers: 360741, 360750, 360751, 360755, 360755 Rev B, 360756, and 360757.

This recall has been terminated (originally issued August 30, 2016).

Company
Ergosafe Products, Llc (Dba) Prism Medical
Recall Initiated
August 30, 2016
Terminated
July 18, 2018
Recall Number
Z-0728-2017
Quantity
not greater than 17,964 units
Firm Location
Maryland Heights, MO
Official Source
View on FDA website ↗

Reason for Recall

The black plastic puck on the carry bar is breaking resulting in potential for patient harm.

Distribution

Distributed US nationwide and to Canada, Japan, Australia, Thailand, Great Britain, and South Korea

Lot / Code Info

There are no lot numbers, serial numbers, or UPC codes, only the model numbers.  Model numbers: 360741, 360750, 360751, 360755, 360755 Rev B, 360756, and 360757.

Root Cause

Device Design

Action Taken

The recalling firm sent field safety notices to their customers by mail on August 30, 2016. The notice instructed customers to inspect their carry bar to see if it used a black plastic puck. If it did, they were instructed to contact the recalling firm to receive a metal insert to install. If the puck is cracked or damaged, they can request a new, replacement carry bar at a discounted price. For further questions, please call (866) 891-6502.

More recalls by Ergosafe Products, Llc (Dba) Prism Medical

View all recalls by this company →