Carry bar which is an accessory to a lift used for patients in hospitals and other facilities. Recall impacts the following model numbers: 360741, 360750, 360751, 360755, 360755 Rev B, 360756, and 360757.
This recall has been terminated (originally issued August 30, 2016).
- Recall Initiated
- August 30, 2016
- Terminated
- July 18, 2018
- Recall Number
- Z-0728-2017
- Quantity
- not greater than 17,964 units
- Firm Location
- Maryland Heights, MO
- Official Source
- View on FDA website ↗
Reason for Recall
The black plastic puck on the carry bar is breaking resulting in potential for patient harm.
Distribution
Distributed US nationwide and to Canada, Japan, Australia, Thailand, Great Britain, and South Korea
Lot / Code Info
There are no lot numbers, serial numbers, or UPC codes, only the model numbers. Model numbers: 360741, 360750, 360751, 360755, 360755 Rev B, 360756, and 360757.
Root Cause
Device Design
Action Taken
The recalling firm sent field safety notices to their customers by mail on August 30, 2016. The notice instructed customers to inspect their carry bar to see if it used a black plastic puck. If it did, they were instructed to contact the recalling firm to receive a metal insert to install. If the puck is cracked or damaged, they can request a new, replacement carry bar at a discounted price. For further questions, please call (866) 891-6502.