A-series lift - A625 Manual w/Load Cell & Wi-Fi, Omni Part No. 324124. Prism Medical.
This recall has been terminated (originally issued February 28, 2017).
- Recall Initiated
- January 30, 2017
- Posted
- February 28, 2017
- Terminated
- December 4, 2017
- Recall Number
- Z-1405-2017
- Quantity
- 21
- Firm Location
- Maryland Heights, MO
- Official Source
- View on FDA website ↗
Reason for Recall
During internal cycle test at max load (625 lbs.) 2 units had straps break prematurely.
Distribution
Product was distributed to MT and NV; one VA consignee.
Lot / Code Info
A6MC000110 A6M0001010 A6M0001011 A6M0001012 A6M0001013 A6M0001014 A6M0001015 A6M0001016 A6M0001017 A6M0001018 A6M0001019 A6M0001020 A6M0001021 A6M0001025 A6M0001026 A6M0001029 A6M0001022 A6M0001023 A6M0001024 A6M0001027 A6M0001028
Root Cause
Device Design
Action Taken
The recalling firm called each customer on Jan 30, 2017 to review the recall and schedule a time for the firm to come to their facility to personally swap the recalled product for a new device. The FSNs were sent on Jan 31, 2017.