Prism Medical P-440 Portable Ceiling Lift; PN 303070 and 303070R (refurbished); Patient lift with weight specification of 440 lbs (200 kg). Product Usage: The P-440 lift is a lifting aid used by health care professionals and those providing care in the home to lift, position, and transfer clients or a disabled family member. The ceiling lift makes it possible to move mobility impaired individuals with minimal strain or risk to the caregiver, while providing complete safety, dignity, and comfort for the client or family member.
This recall has been terminated (originally issued July 27, 2015).
- Recall Initiated
- June 24, 2015
- Posted
- July 27, 2015
- Terminated
- April 3, 2017
- Recall Number
- Z-2269-2015
- Quantity
- 7105 lifts
- Firm Location
- Maryland Heights, MO
- Official Source
- View on FDA website ↗
Reason for Recall
There is a potential the sling loops may not stay attached to the carry bar hooks.
Distribution
Worldwide Distribution - US Nationwide and the countries of:Canada, Great Britain, Australia, South Korea, and Japan.
Lot / Code Info
All serial numbers.
Root Cause
Other
Action Taken
Prism Medical sent a Field Safety Notice to affected customers via certified first class US mail. The notice identified the affected product, problem and actions to be taken. For questions contact your local Prism Medical dealer or Prism Medical Customer Service Representative 1-800-891-6502
More recalls by Ergosafe Products, Llc (Dba) Prism Medical
Carry bar which is an accessory to a lift used for patients in hospitals and other facilities. Reca...