Synthes Ti Button Plate 7 Holes, Fastener, Fixation, nondegradable, Soft Tissue

This recall has been terminated (originally issued July 7, 2016).

Company
Synthes (Usa) Products Llc
Recall Initiated
May 31, 2016
Posted
July 7, 2016
Terminated
April 25, 2017
Recall Number
Z-2138-2016
Quantity
6,255 units
Firm Location
West Chester, PA
Official Source
View on FDA website ↗

Reason for Recall

DePuy Synthes is voluntarily initiating a recall due to affected devices being distributed and or labeled with MR Safety information that does not meet the current ASTM standard.

Distribution

USA Distribution to the states of : AK,AL,AR,AZ,CA,CH,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,ON,OR,PA,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI, & WV

Lot / Code Info

Lot Numbers: 1749148, 1896349, 3057585, 3126474, 3277587, 3527962, 3666979, 3723149, 3754359, 3754360, 3800237, 3800239, 3825498, 4182364, 4182365, 4184238, 4184239, 4184240, 4184241, 4186156, 4186158, 4186159, 4201866, 4201867, 4201868, 4205288, 4872583, 4894156, 4894157, 4931671, 4931672, 4931673, 4952212, 4992237, 4992836, 5004334, 5083067, 5106349, 5115240, 5115241, 5127039, 5153048, 5153049, 5328496, 5337095, 5339819, 5344347, 5369177, 5391267, 5404749, 5450581, 5485082, 5510560, 5510561, 5510562, 5510563, 5510564, 5510565, 5583188, 7554579, 7554580, 7559611, 7559614, 7571578, 7571579, 7580696, 7583500, 7596250, 7780217, 7784418, 7784419, 7875460, 7923388, 7949747, 7992504, 8024598, 8060315. 8155523, 8203696, 8226175, 8238466, 8292917,8312365,8348012,8377148,8504955,8612171,8688476,8701156,8711152,8743078,8743084 8756426, 8756427, 8834945,8840904,8948836,8948838,9079710,9079712,9079716,9172619,9172621,9237247 9361663, 9361665, 9432606, 9432608, 9494467, 9494469, 9608863, 9647283, 9665152, 9671639, 9715870, 9720837, 9720840, 9783567, 9783570,. 9788216, 9801777, 9810012, 9823178, 9829378, 9836403, 9836404, 9836405, 9896775, 9913152 Part Numbers: 482.823, Device Listing Number: D194880

Root Cause

Under Investigation by firm

Action Taken

Synthes mailed to customers (Hospital Personnel and Sales Consultants) an URGENT FIELD SAFETY NOTIFICATION letter informing them the insert they received, which contained incorrect MR Safety Information, does not meet the current ASTM standards. Customers were asked to review the package insert provided with the notification letter, discard the outdated package insert, complete the attached verification Section (page 3 of this notification) and send a copy of the completed Verification Section by Fax: 844-782-5568 or scan/email: Synthes7923@stericycle.com. For further questions call (610) 719-6500.