Synthes Ti Button Plate 7 Holes, Fastener, Fixation, nondegradable, Soft Tissue
This recall has been terminated (originally issued July 7, 2016).
- Company
- Synthes (Usa) Products Llc
- Recall Initiated
- May 31, 2016
- Posted
- July 7, 2016
- Terminated
- April 25, 2017
- Recall Number
- Z-2138-2016
- Quantity
- 6,255 units
- Firm Location
- West Chester, PA
- Official Source
- View on FDA website ↗
Reason for Recall
DePuy Synthes is voluntarily initiating a recall due to affected devices being distributed and or labeled with MR Safety information that does not meet the current ASTM standard.
Distribution
USA Distribution to the states of : AK,AL,AR,AZ,CA,CH,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,ON,OR,PA,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI, & WV
Lot / Code Info
Lot Numbers: 1749148, 1896349, 3057585, 3126474, 3277587, 3527962, 3666979, 3723149, 3754359, 3754360, 3800237, 3800239, 3825498, 4182364, 4182365, 4184238, 4184239, 4184240, 4184241, 4186156, 4186158, 4186159, 4201866, 4201867, 4201868, 4205288, 4872583, 4894156, 4894157, 4931671, 4931672, 4931673, 4952212, 4992237, 4992836, 5004334, 5083067, 5106349, 5115240, 5115241, 5127039, 5153048, 5153049, 5328496, 5337095, 5339819, 5344347, 5369177, 5391267, 5404749, 5450581, 5485082, 5510560, 5510561, 5510562, 5510563, 5510564, 5510565, 5583188, 7554579, 7554580, 7559611, 7559614, 7571578, 7571579, 7580696, 7583500, 7596250, 7780217, 7784418, 7784419, 7875460, 7923388, 7949747, 7992504, 8024598, 8060315. 8155523, 8203696, 8226175, 8238466, 8292917,8312365,8348012,8377148,8504955,8612171,8688476,8701156,8711152,8743078,8743084 8756426, 8756427, 8834945,8840904,8948836,8948838,9079710,9079712,9079716,9172619,9172621,9237247 9361663, 9361665, 9432606, 9432608, 9494467, 9494469, 9608863, 9647283, 9665152, 9671639, 9715870, 9720837, 9720840, 9783567, 9783570,. 9788216, 9801777, 9810012, 9823178, 9829378, 9836403, 9836404, 9836405, 9896775, 9913152 Part Numbers: 482.823, Device Listing Number: D194880
Root Cause
Under Investigation by firm
Action Taken
Synthes mailed to customers (Hospital Personnel and Sales Consultants) an URGENT FIELD SAFETY NOTIFICATION letter informing them the insert they received, which contained incorrect MR Safety Information, does not meet the current ASTM standards. Customers were asked to review the package insert provided with the notification letter, discard the outdated package insert, complete the attached verification Section (page 3 of this notification) and send a copy of the completed Verification Section by Fax: 844-782-5568 or scan/email: Synthes7923@stericycle.com. For further questions call (610) 719-6500.