RecallDepth

LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.

Company
Synthes (Usa) Products Llc
Recall Initiated
September 26, 2025
Posted
October 21, 2025
Recall Number
Z-0229-2026
Quantity
122 units
Firm Location
West Chester, PA

Reason for Recall

Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MN, MO, MS, NJ, NM, NY, OH, TN, TX and the country of Canada.

Lot / Code Info

Product Code: 04.045.342TS. GTIN: 10886982301390. Lot Number: 31756P9. Expiration Date: 08/31/2025

Root Cause

Process control

Action Taken

DePuy Synthes notified consignees via email and mailed letter on 09/26/2025. Consignees were instructed to examine inventory and quarantine affected units, arrange for the return of any affected units, complete and return the Business Response Form, and place the notification in a visible area for awareness. If units were further distributed within your facility or to customers, please notify those customers and personnel.

More recalls by Synthes (Usa) Products Llc