RecallDepth

STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010

Company
Synthes (Usa) Products Llc
Recall Initiated
April 25, 2024
Posted
May 28, 2024
Recall Number
Z-1916-2024
Quantity
261 units
Firm Location
West Chester, PA

Reason for Recall

Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Lot / Code Info

UDI-DI: 10886982075192 Lot Numbers: 97P6250, 97P6251, 97P6252

Root Cause

Process change control

Action Taken

DePuy Synthes issued Urgent Medical Device Recall (Removal) Letter on 4/26/24. Letter states reason for recall, health risk and action to take: Examine your inventory to determine if you have the subject product lots and quarantine them immediately. DO NOT USE THE SUBJECT PRODUCTS. 2. Contact your DePuy Synthes Sales Consultant or contact the customer support services at ra-dpyuscsspecial@ its.jnj.com to coordinate the return/credits of the subject product lots. 3. Review, complete, sign, and return the attached Business Response Form (page 3 of this letter) to OneMDField- Actions@its.jnj.com within three (3) business days of receipt of this notification. Please include in the email subject: FA 2363331. 4. Please complete the attached Business Response Form even if you do not have the subject product lots on hand. 5. Forward this notice to anyone in your facility that needs to be informed (e.g., those who manage, transport, store, stock, or use the subject products). 6. If any of the subject product lots have been forwarded to another facility, contact that facility and provide them with this notice. 7. Post a copy of this notice in a visible area for awareness and keep a copy for your records. If you have any questions, please contact your local DePuy Synthes Sales Consultant. For Medical Information request, please visit our website: https://www.jnjmedicaldevices.com/mir.

More recalls by Synthes (Usa) Products Llc

View all recalls by this company →