STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010
- Company
- Synthes (Usa) Products Llc
- Recall Initiated
- April 25, 2024
- Posted
- May 28, 2024
- Recall Number
- Z-1916-2024
- Quantity
- 261 units
- Firm Location
- West Chester, PA
Reason for Recall
Stardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Lot / Code Info
UDI-DI: 10886982075192 Lot Numbers: 97P6250, 97P6251, 97P6252
Root Cause
Process change control
Action Taken
DePuy Synthes issued Urgent Medical Device Recall (Removal) Letter on 4/26/24. Letter states reason for recall, health risk and action to take: Examine your inventory to determine if you have the subject product lots and quarantine them immediately. DO NOT USE THE SUBJECT PRODUCTS. 2. Contact your DePuy Synthes Sales Consultant or contact the customer support services at ra-dpyuscsspecial@ its.jnj.com to coordinate the return/credits of the subject product lots. 3. Review, complete, sign, and return the attached Business Response Form (page 3 of this letter) to OneMDField- Actions@its.jnj.com within three (3) business days of receipt of this notification. Please include in the email subject: FA 2363331. 4. Please complete the attached Business Response Form even if you do not have the subject product lots on hand. 5. Forward this notice to anyone in your facility that needs to be informed (e.g., those who manage, transport, store, stock, or use the subject products). 6. If any of the subject product lots have been forwarded to another facility, contact that facility and provide them with this notice. 7. Post a copy of this notice in a visible area for awareness and keep a copy for your records. If you have any questions, please contact your local DePuy Synthes Sales Consultant. For Medical Information request, please visit our website: https://www.jnjmedicaldevices.com/mir.