Hitachi AIRIS II MRI System

This recall has been terminated (originally issued June 14, 2006).

Company
Hitachi Medical Systems America Inc
Recall Initiated
April 25, 2006
Posted
June 14, 2006
Terminated
October 29, 2008
Recall Number
Z-1115-06
Quantity
248
Firm Location
Twinsburg, OH
Official Source
View on FDA website ↗

Reason for Recall

Improper Assembly-The AIRIS/AIRIS II MRI system has the potential for the input wiring of a power transformer to overheat due to damaged wiring or possible loose connections.

Distribution

The devices were distributed to consignees located throughout the United States.

Lot / Code Info

Serial Numbers: C322- to C546; and C746 to C772

Root Cause

Other

Action Taken

The recalling firm plans to inspect the transformer connections for damaged wiring and check the connections for tightness and tighten as necessary. The inspections and possible corrections began on 4/25/2006 and are in the process of being conducted by the firm''s Field Service Engineers. The firm anticipates trhat these actions will be completed by 6/25/2006.

More recalls by Hitachi Medical Systems America Inc