ADVIA 1800 System Software V2.01. Performs assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse testing and therapeutic drug monitoring.

This recall has been terminated (originally issued November 18, 2013).

Company
Siemens Healthcare Diagnostics
Recall Initiated
September 3, 2013
Posted
November 18, 2013
Terminated
March 7, 2017
Recall Number
Z-0345-2014
Quantity
1366 units
Firm Location
Tarrytown, NY
Official Source
View on FDA website ↗

Reason for Recall

Siemens Healthcare Diagnostics issued a Urgent Medical Device Correction due to Ration Parameters Flagging Behavior for All Software Versions utilized on ADVIA 1200, 1650, 1800 or 2400 Chemistry Systems.

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Sweden, Switzerland, Taiwan, Turkey, United Kingdom, and Vietnam.

Lot / Code Info

Product Code: 10639244

Root Cause

Software design

Action Taken

On September 3, 2013, Siemens Healthcare Diagnostics issued an Urgent Field Safety Notice for the ADVIA 1200, 1650, 1800 and 2400 Chemistry Systems due to a software issue. The software issue has caused the Chemistry Systems to generate reports without error flags. Notices were sent to domestic customers via U.S. and outside the US the distribution is determined at the country level. The customers were instructed to contact Siemens Healthcare Diagnostics Technical Solution Center if they have technical questions regarding this issue at 877-229-3711 (prompt 18).

More recalls by Siemens Healthcare Diagnostics