IMMULITE /IMMULITE 1000 Systems COR Cortisol, REF/Catalog Number LKC01, SMN 10381388, IVD. For in vitro diagnostic use with the IMMULITE 1000 Systems Analyzers for the quantitative measurement measurement of cortisol (hydrocortisone, Compound F) in serum, as an aid in the clinical assessment of adrenal status.

This recall has been terminated (originally issued October 23, 2014).

Company
Siemens Healthcare Diagnostics
Recall Initiated
September 19, 2014
Posted
October 23, 2014
Terminated
November 18, 2015
Recall Number
Z-0122-2015
Quantity
US: 772 units, Canada: 43 units
Firm Location
Tarrytown, NY
Official Source
View on FDA website ↗

Reason for Recall

Customer complaints were received for positive bias reported on the controls used on the IMMULITE /IMMULITE 1000 on the Cortisol Assay reagent lots IMMULITE /IMMULITE 1000 (LKCO1) Lots 381,382 and 383 . Subsequently, the firm confirmed a positive bias on the IMMULITE/IMMULITE 1000 (LKCO1) Lots 381, 382 and 383.

Distribution

Worldwide Distribution: US (nationwide) and internationally to Canada.

Lot / Code Info

Lots: 381 (exp. 10/31/2014), 382 (exp. 10/31/2014) and 383 (exp. 11/30/2014)

Root Cause

Nonconforming Material/Component

Action Taken

US Customers were sent an Urgent Medical Device Recall Letter (UMDR2014-09-19, dated 9/19/14) via FedEx on 9/19/14. Outside US Customers were sent an Urgent Field Safety Notice Letter. The letters inform the consignees of the recall of the IMMULITE/IMMULITE 1000 Cortisol (LKCO1) assay, kit lots 381, 382 and 383. Customers were asked to take the following actions: - Please review this letter with your Medical Director. - Discontinue use of and discard the IMMULITE/IMMULITE 1000 Cortisol kit lots listed in the letter. - Review your inventory of these products to determine your laboratorys replacement needs and to provide information to Siemens for reporting to the authorities. --- Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Field Safety Notice letter within thirty (30) days.