BC Thrombin Reagent Kit Lot Number 46751; UDI - 008427680131024675120180908 and BC Thrombin Reagent Kit Lot Number 47184; UDI -008427680131024718420190221; BC Thrombin Reagent; Test, Thrombin Time Product Usage: For in vitro diagnostic use. Reagent for the quantitative determination of the thrombin time in citrated human plasma.
This recall has been terminated (originally issued October 17, 2017).
- Company
- Siemens Healthcare Diagnostics
- Recall Initiated
- October 17, 2017
- Terminated
- November 4, 2019
- Recall Number
- Z-0827-2018
- Quantity
- 7631 units
- Firm Location
- Tarrytown, NY
- Official Source
- View on FDA website ↗
Reason for Recall
The BC Thrombin Reagent kit lot 46751 (contains Thrombin Reagent lot 517468) and kit lot 47184 (contains Thrombin Reagent lot 517469) produce unexpected prolonged Thrombin Time (TT) results for expected normal samples, and may recover above the upper limit of normal (< 21 sec.) as stated within the IFU (Instruction For Use).
Distribution
Worldwide distribution. US nationwide, Austria, Belgium, Bosnia-Herzegovina, Croatia, Finland, France (incl. Guadeloupe, Martinique, Reunion, French Guinea overseas departments), Germany, Greece, Hungary, Italy, Kazakhstan, Macedonia, Latvia, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, Spain, Sweden, Turkey, Argentina, Brazil, El Salvador, Mexico, Peru, Uruguay, Canada, and Australia.
Lot / Code Info
Manufactured: BC Thrombin Reagent Kit Lot 46751: 3673 kits BC Thrombin Reagent Kit Lot 47184: 3958 kits Total kits Distributed to Customers Domestic and Foreign: BC Thrombin Reagent Kit Lot 46751: 3549 kits BC Thrombin Reagent Kit Lot 47184: 2580 kits Product Code/ Lot# Date Manufactured Date Expired 46751 2016-11-01 2018-09-08 47184 2017-03-20 2019-02-21
Root Cause
Nonconforming Material/Component
Action Taken
Siemens sent an Urgent Field Safety Notice letter dated October 2017 to affected customer. The letter identified the affected product, problem and actions to be taken by the customer. Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter and the Urgent Field Corrective Action letter within 30 days. For questions contact your Siemens Customers Care Center or local Siemens technical support representative.