BC Thrombin Reagent Kit Lot Number 46751; UDI - 008427680131024675120180908 and BC Thrombin Reagent Kit Lot Number 47184; UDI -008427680131024718420190221; BC Thrombin Reagent; Test, Thrombin Time Product Usage: For in vitro diagnostic use. Reagent for the quantitative determination of the thrombin time in citrated human plasma.

This recall has been terminated (originally issued October 17, 2017).

Company
Siemens Healthcare Diagnostics
Recall Initiated
October 17, 2017
Terminated
November 4, 2019
Recall Number
Z-0827-2018
Quantity
7631 units
Firm Location
Tarrytown, NY
Official Source
View on FDA website ↗

Reason for Recall

The BC Thrombin Reagent kit lot 46751 (contains Thrombin Reagent lot 517468) and kit lot 47184 (contains Thrombin Reagent lot 517469) produce unexpected prolonged Thrombin Time (TT) results for expected normal samples, and may recover above the upper limit of normal (< 21 sec.) as stated within the IFU (Instruction For Use).

Distribution

Worldwide distribution. US nationwide, Austria, Belgium, Bosnia-Herzegovina, Croatia, Finland, France (incl. Guadeloupe, Martinique, Reunion, French Guinea overseas departments), Germany, Greece, Hungary, Italy, Kazakhstan, Macedonia, Latvia, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, Spain, Sweden, Turkey, Argentina, Brazil, El Salvador, Mexico, Peru, Uruguay, Canada, and Australia.

Lot / Code Info

Manufactured: BC Thrombin Reagent Kit Lot 46751: 3673 kits BC Thrombin Reagent Kit Lot 47184: 3958 kits  Total kits Distributed to Customers Domestic and Foreign:  BC Thrombin Reagent Kit Lot 46751: 3549 kits BC Thrombin Reagent Kit Lot 47184: 2580 kits  Product Code/ Lot# Date Manufactured Date Expired 46751 2016-11-01 2018-09-08 47184 2017-03-20 2019-02-21

Root Cause

Nonconforming Material/Component

Action Taken

Siemens sent an Urgent Field Safety Notice letter dated October 2017 to affected customer. The letter identified the affected product, problem and actions to be taken by the customer. Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter and the Urgent Field Corrective Action letter within 30 days. For questions contact your Siemens Customers Care Center or local Siemens technical support representative.