RecallDepth

Xoran MiniCAT CT Scanner ID ENT00001

This recall has been terminated (originally issued December 14, 2009).

Company
Xoran Technologies, Inc.
Recall Initiated
December 14, 2009
Terminated
September 11, 2014
Recall Number
Z-0041-2014
Quantity
320
Firm Location
Ann Arbor, MI
Official Source
View on FDA website ↗

Reason for Recall

Failure of the gantry to rotate. The complaint alleged that the scanner would emit radiation but would not rotate.

Distribution

Worldwide Distribution - US nationwide

Lot / Code Info

minicat

Root Cause

Radiation Control for Health and Safety Act

Action Taken

The service report in response to this complaint stated that the Head select relay board, chip and fan were replaced. The machine (ENTOOOO I) was subsequently tested. All tests were successful.