Xoran MiniCAT CT Scanner ID ENT00001
This recall has been terminated (originally issued December 14, 2009).
- Company
- Xoran Technologies, Inc.
- Recall Initiated
- December 14, 2009
- Terminated
- September 11, 2014
- Recall Number
- Z-0041-2014
- Quantity
- 320
- Firm Location
- Ann Arbor, MI
- Official Source
- View on FDA website ↗
Reason for Recall
Failure of the gantry to rotate. The complaint alleged that the scanner would emit radiation but would not rotate.
Distribution
Worldwide Distribution - US nationwide
Lot / Code Info
minicat
Root Cause
Radiation Control for Health and Safety Act
Action Taken
The service report in response to this complaint stated that the Head select relay board, chip and fan were replaced. The machine (ENTOOOO I) was subsequently tested. All tests were successful.