RecallDepth

ADVIA Centaur CP System, Automated Immunoassay Analyzer, Part /Catalogue Number 086-A001.

This recall has been terminated (originally issued September 23, 2006).

Company
Siemens Healthcare Diagnostics
Recall Initiated
August 4, 2006
Posted
September 23, 2006
Terminated
July 30, 2012
Recall Number
Z-1516-06
Quantity
Domestic - 101, Foreign - 153
Firm Location
Tarrytown, NY
Official Source
View on FDA website ↗

Reason for Recall

Sample/Patient mis-identification (software defect)-the system can associate a test result and sample identification (SID) with an incorrect patient name when the Patient Demographics feature is used and the Patient Identification (PID) field is left blank

Distribution

Worldwide

Lot / Code Info

All ADVIA Centaur CP systems using software versions 3.1 and 3.2.

Root Cause

Software design

Action Taken

On 8/04/06, a Support Bulletin was e-mailed to all affected Bayer HealthCare LLC Branches worldwide in order to inform them of the issue and to provide them with instructions for this 'field correction' until the new software version is issued. A Customer Bulletin and Confirmation Form (FaxBack) were prepared and sent to the affected Bayer Healthcare LLC Branches on 8/04/06 for communication with the affected customers worldwide.

More recalls by Siemens Healthcare Diagnostics

View all recalls by this company →