AGFA IMPAX HeartStation ECG Management System.

Company
Agfa Corp.
Recall Initiated
May 7, 2009
Posted
November 18, 2009
Terminated
July 27, 2010
Recall Number
Z-0366-2010
Quantity
9 units
Firm Location
Greenville, SC

Reason for Recall

HeartStation does not refresh content changes to the PDF reports that are printed emailed or faxed from the Index page. This may lead to a discrepancy in patient demographics between what is displayed in the HeartStation Client interface and the report delivered to the consumer. Questions are being directed to the local Agfa representative; the contact information was included in the letter.

Distribution

CA, KY, MA, NY, OH, SC, Belgium, Italy and Singapore

Lot / Code Info

Model number HeartStation 1.1 60+00091179 to 60+0091186; HeartStation Core Server 60+00091211; CV HeartStation EZKT7000, L9NBN000, EZKC000, L9NAL000, EZKWE000 and L9M9G000.

Root Cause

Other

Action Taken

An Urgent Safety Notice was sent via FED-EX to all the sites with the affected units (a total of 6) on May 7, 2009. The three foreign accounts were notified by letter on May 14, 2009. The letter describes the potential issue and mitigation. Acknowledgment via FAX-Back that the information was received and understood was requested from the sites. Customers who use the print, email, fax functionality of this product were verbally notified of the defect and they have been advised of a workaround. In the notice letter, Agfa HealthCare is also providing customers with the recommended workaround. The workaround is to only print from the Viewer screen or to print the ECG once confirmed. The Viewer screen, however, does not allow the user to print batches of reports as does the Index screen. Questions about this matter are being directed to the local Agfa HealthCare representative.

More recalls by Agfa Corp.

View all recalls by this company →