CELL-Dyn Ruby Hematology Analyzer, List Number 08H67-01 The CELL-DYN Ruby is a multi-parameter, automated hematology analyzer designed for in vitro diagnostic use in clinical laboratories. The instrument has two modes of operation: the Autoloader mode that aspirates samples from closed collection tubes and the Open Tube mode that aspirates samples from open collection tubes.
- Company
- Abbott Laboratories
- Recall Initiated
- November 17, 2008
- Posted
- April 10, 2009
- Terminated
- November 27, 2010
- Recall Number
- Z-1064-2009
- Quantity
- 899 units
- Firm Location
- Santa Clara, CA
Reason for Recall
Carryover failures in software revisions 1.0ML and 2.0ML result in elevated PLT background count.
Distribution
Worldwide Distribution
Lot / Code Info
Serial No. 34001BG - 34965BG
Root Cause
Other
Action Taken
Recall initiated on 11/17/2008. A Product Correction letter and customer reply form is sent to all currently active CELL-DYN Ruby customers. The customers are asked to keep the communication with their CELL-DYN Ruby System Operator's Manual and to note that the information listed in the table replaces the carryover specification provided in all applications of the manual. Questions from customers in the US should be directed to Customer Support at 1-877-4ABBOTT. Customers outside the US should contact their local hematology customer support representative.
More recalls by Abbott Laboratories
Brand Name: Alinity hq Analyzer
Product Name: Alinity hq Analyzer
Model/Catalog Number: 09P68-01
Sof...
Nov 12, 2025
Alinity hq Analyzer, REF: 09P68-01
Aug 16, 2024
Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number...
May 9, 2019