Monaco Radiation Treatment Planning Workstation. Monaco Release 2.04.00 and above. Used to create treatment plans for any cancer patient for who external beam intensity modulated radiation therapy (IMRT) has been prescribed.

This recall has been terminated (originally issued June 15, 2011).

Company
Computerized Medical Systems Inc
Recall Initiated
October 27, 2010
Posted
June 15, 2011
Terminated
April 2, 2014
Recall Number
Z-2569-2011
Quantity
11
Firm Location
Maryland Heights, MO
Official Source
View on FDA website ↗

Reason for Recall

Monaco: In XiO IMRT (Dynamic MLC Delivery) software, the radiation beam angle is set to zero. If different beam angles are selected via Monaco software, the beam angles will remain at zero.

Distribution

Worldwide Distribution - USA including CA, IL, MN, NM, OH, NY, TX, and WA and the Netherlands.

Lot / Code Info

Release 2.04 and higher

Root Cause

Software design

Action Taken

Computerized Medical Systems, Inc. sent a User Notice to the affected consignees on October 27, 2010, explaining the reason for correction, the clinical impact, and the workaround method. (VERIFY USE NOTICE DATE) A return postcard was included for the customer to confirm receipt of the User Notice. The firm intends for this issue to be corrected in Monaco Release 3.00 in March 2011.