Agfa Impax R6.3 DB Server. Impax Systems are Picture Archiving and Communication systems (PACs). PAC systems are used in the acceptance, transfer, display, storage and digital processing of medical images. Primary users are radiologists, mammographers, orthopaedists and other trained medical staff.

Company
Agfa Corp.
Recall Initiated
July 16, 2009
Posted
September 8, 2009
Terminated
April 28, 2011
Recall Number
Z-1887-2009
Quantity
One unit
Firm Location
Greenville, SC

Reason for Recall

Measurements are inaccurate on multiple workstations.

Distribution

Nationwide Distribution -- IL

Lot / Code Info

Product Version 6.3.1 SU10.

Root Cause

Software design

Action Taken

Agfa Healthcare issued an "URGENT FIELD SAFETY NOTICE" dated July 16, 2009. The notice informed of the problem and provided details to identify potentially compromised data that could result in measurement inaccuracies. A feedback form was provided to be completed and returned to the firm. For further information, contact Agfa Healthcare at 1-773-592-4769.