RecallDepth

NX 2008 Central Monitoring System (CMS). Product code: E2FB6 Agfa's Computed Radiography Systems with NX Workstations are intended for use in providing diagnostic quality images to aid the physician with diagnosis.

Company
Agfa Corp.
Recall Initiated
June 22, 2009
Posted
September 2, 2009
Terminated
July 22, 2010
Recall Number
Z-1889-2009
Quantity
97 units
Firm Location
Greenville, SC

Reason for Recall

After opening and closing a study on a Central Monitoring System while a study is still open on an in-room NX system, a problem of image mix-up can occur on the in-room NX system.

Distribution

Worldwide - IL, PA, TX, Canada, Europe, Asia Pacific and Latin America

Lot / Code Info

NX Software versions: 2.0. 7000, 3.0.7000, 2.0.8000 and 3.0.8000.

Root Cause

Software design

Action Taken

An "Urgent Safety Notice" was sent to all potentially affected sites on/about June 22, 2009 to make them aware of the issue. A FAX Back form is requested with each letter to ensure someone at the account acknowledges the information was received and understood. If a FAX Back form is not received, the account will be contacted by telephone. A software service update to solve this issue is expected to be released by August 15, 2009. A Mandatory Service Bulletin will be deployed to correct the problem in the software. If you have questions, please contact Agfa Healthcare at 1-800-581-2432 prompts 1, 3 and then 2 for CR Support.

More recalls by Agfa Corp.

View all recalls by this company →