Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

Company
Fresenius Kabi Usa, Llc
Recall Initiated
January 10, 2025
Posted
February 4, 2025
Recall Number
Z-1019-2025
Quantity
23 systems
Firm Location
North Andover, MA

Reason for Recall

Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.

Distribution

US distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI.

Lot / Code Info

UDI-DI: 00811505030122. Software 5.9.2 and prior

Root Cause

Software Design Change

Action Taken

Fresenius Kabi began notifying consignees on 01/10/2024 via emailed letter. Consignees were instructed to check if your device contains the affected software, review and become familiar with the temporary Risk Mitigation actions for each anomaly outlined in the letter and notify customers/personnel if the product was transferred or further distributed. Additionally, to facilitate the installation of the pump software, LVP SW 5.10, your facility must install the new Ivenix Infusion Management System (IMS) software version (5.2) on your IMS server and install the new software version (5.10) for the Ivenix Large Volume Pump (LVP). Please post this letter in areas that affected devices are stored and used such that users of this device are aware of the issues and recommended risk mitigation measures. Lastly, consignees were requested to complete and return the Customer Reply Form.