Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Software for the controlled administration of fluids to patients.

Company
Fresenius Kabi Usa, Llc
Recall Initiated
November 14, 2025
Posted
December 16, 2025
Recall Number
Z-0885-2026
Quantity
30 units
Firm Location
North Andover, MA

Reason for Recall

Software version 5.10.1 and earlier contain anomalies that have the potential to cause serious patient harm or death.

Distribution

US (Domestic) distribution: CA, CO, FL, GA, ID, IL, MD, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA & WI.

Lot / Code Info

Software Versions 5.10.1 and prior; UDI: 00811505030122.

Root Cause

Software design

Action Taken

Consignees were emailed an "URGENT - Medical Device Field Correction" notice dated 11/14/25. The notice instructs consignees to review and become familiar with the provided notification's risk mitigation actions; the notice is to be forwarded if affected units were further distributed and users are to be notified. Consignees are to facilitate the installation of the new Ivenix Infusion Management System (version 5.2.2) and Large Volume Pump (5.10.2) software versions utilizing the provided instructions. Consignees are to note that the LVP will not be available for use during the update. Completed Customer Reply Forms may be sent via email to fkmd.fieldactions@fresenius-kabi.com. Customer questions can be directed to Customer Support via email at Ivenix_support@fresenius-kabi.com or by phone at 1-855-354-6387.